Share Email Print
cover

Proceedings Paper

Component validation of direct diode 488nm lasers in BD Accuri C6 flow cytometers
Author(s): Wei P. Chen; Ningyi D. Luo
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

The 488nm laser is the most important excitation light source of flow cytometry. The indirect diode (frequency-doubled diode) 488nm lasers are used in the excitation of Becton Dickinson (BD) AccuriTM C6. For using cost effective lasers, we have validated direct diode 488nm lasers as the replacement component of frequency-doubled diode laser. BD Bioscience issued the protocols to cover wavelength, power, noise, and polarization at the operation temperature range of cytometer. Pavilion Integration Corporation (PIC) tested 6 samples as the component validation of direct diode 488nm lasers based on the protocols from BD Biosciences. BD Bioscience also tested one of laser samples to further validate the test results of power, noise, and polarization from PIC.

Paper Details

Date Published: 18 March 2016
PDF: 5 pages
Proc. SPIE 9700, Design and Quality for Biomedical Technologies IX, 97000E (18 March 2016); doi: 10.1117/12.2213432
Show Author Affiliations
Wei P. Chen, Becton Dickinson Biosciences (United States)
Ningyi D. Luo, Pavilion Integration Corp. (United States)


Published in SPIE Proceedings Vol. 9700:
Design and Quality for Biomedical Technologies IX
Ramesh Raghavachari; Rongguang Liang, Editor(s)

© SPIE. Terms of Use