A polarimeter for characterization of the instrumental polarization of an imaging UV spectrometer has been designed. In the course of calibrating the polarimeter, the quarter-wave retarders were observed to possess polarization properties other than pure linear retardance. This forced the development of a complex procedure for the calibration of the retarders and a new generalized approach to polarimetric analysis.

Paper Details

Date Published: 1 October 1990
PDF: 11 pages
Proc. SPIE 1317, Polarimetry: Radar, Infrared, Visible, Ultraviolet, and X-Ray, (1 October 1990); doi: 10.1117/12.22074
Show Author Affiliations
M. Frank Morgan, Univ. of Alabama in Huntsville (United States)
Russell A. Chipman, Univ. of Alabama in Huntsville (United States)
Douglas G. Torr, Univ. of Alabama in Huntsville (United States)