Share Email Print
cover

Proceedings Paper

Research on calibration technique for a high-precision optical angle gauge
Author(s): Jiaming Lin; Genrui Cao; Datu Su
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

Optical angle gauge is one of the measuring instruments standardized by the National Metrological System of Chian for verification of plane angle. In this paper a new method for verifying the optical angle gauge is described and a related photoelectric goniometer developed by ourself is introduced. As a datum instrument for measuring deviation angle of optical wedge, its maximum error is less than 0.1 inch. In addition, some examples of its application for calibrating indication error of various angle measuring devices are presented, and its distinctive advantages, especially in calibrating 2D autocollimators, are demonstrated.

Paper Details

Date Published: 8 September 1995
PDF: 8 pages
Proc. SPIE 2536, Optical Manufacturing and Testing, (8 September 1995); doi: 10.1117/12.218459
Show Author Affiliations
Jiaming Lin, Beijing Institute of Technology (China)
Genrui Cao, Beijing Institute of Technology (China)
Datu Su, Beijing Institute of Technology (China)


Published in SPIE Proceedings Vol. 2536:
Optical Manufacturing and Testing
Victor J. Doherty; H. Philip Stahl, Editor(s)

© SPIE. Terms of Use