In the paper, a computer system for measurement, visualization and analysis of intracranial pressure (ICP), medium arterial pressure and cerebral perfusion pressure in one, two, three or four patients simultaneously has been presented. A structure of pneumatic compensatory sensor for intracranial pressure, and a stand for static properties of the sensors testing has been discussed. Conclusions resulting from the period of using the monitoring system with ICP pneumatic sensors have been formulated.

Paper Details

Date Published: 30 June 1995
PDF: 7 pages
Proc. SPIE 2634, Optoelectronic and Electronic Sensors, (30 June 1995); doi: 10.1117/12.213152
Show Author Affiliations
Henryk M. Juniewicz, Wroclaw Technical Univ. (Poland)
Miroslaw Werszko, Wroclaw Technical Univ. (Poland)