Algorithms for skiascopy measurement automatization
Author(s): Sergejs Fomins; Renārs Trukša; Gunta Krūmiņa
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

Automatic dynamic infrared retinoscope was developed, which allows to run procedure at a much higher rate. Our system uses a USB image sensor with up to 180 Hz refresh rate equipped with a long focus objective and 850 nm infrared light emitting diode as light source. Two servo motors driven by microprocessor control the rotation of semitransparent mirror and motion of retinoscope chassis. Image of eye pupil reflex is captured via software and analyzed along the horizontal plane. Algorithm for automatic accommodative state analysis is developed based on the intensity changes of the fundus reflex.

Paper Details

Date Published: 22 October 2014
PDF: 6 pages
Proc. SPIE 9421, Eighth International Conference on Advanced Optical Materials and Devices (AOMD-8), 94210N (22 October 2014); doi: 10.1117/12.2083950
Show Author Affiliations
Sergejs Fomins, Univ. of Latvia (Latvia)
Renārs Trukša, Univ. of Latvia (Latvia)
Gunta Krūmiņa, Univ. of Latvia (Latvia)