Imaging performance of a large-field kinestatic charge detector for digital radiography
Author(s): Frank A. DiBianca; Sreenivas Devidas; George C. Giakos; Shashidhar Kollipara; Joseph S. Laughter; Azad Mahmud; Senthilkumar Nagarajan; Qian Peng; Carlos Rodriguez; Herbert D. Zeman
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

The initial performance of a digital radiographic system incorporating a large-field (2016- channel) kinestatic charge detector and data acquisition electronics is discussed. The measured modulation transfer function of the system is 20% at 4 cy/mm. The measured detective quantum efficiency is 40 - 60%. These results are comparable with or better than those of current clinical (rare-earth film-screen and storage phosphor) systems. First images from the large-field system are shown and compared with those from commercial systems. Future system improvements in process or in planning are discussed.

Paper Details

Date Published: 8 May 1995
PDF: 12 pages
Proc. SPIE 2432, Medical Imaging 1995: Physics of Medical Imaging, (8 May 1995); doi: 10.1117/12.208359
Show Author Affiliations
Frank A. DiBianca, Univ. of Tennessee/Memphis (United States)
Sreenivas Devidas, Univ. of Tennessee/Memphis (United States)
George C. Giakos, Univ. of Tennessee/Memphis (United States)
Shashidhar Kollipara, Univ. of Tennessee/Memphis (United States)
Joseph S. Laughter, Univ. of Tennessee/Memphis (United States)
Azad Mahmud, Univ. of Tennessee/Memphis (United States)
Senthilkumar Nagarajan, Univ. of Tennessee/Memphis (United States)
Qian Peng, Univ. of Tennessee/Memphis (United States)
Carlos Rodriguez, Univ. of Tennessee/Memphis (United States)
Herbert D. Zeman, Univ. of Tennessee/Memphis (United States)


Published in SPIE Proceedings Vol. 2432:
Medical Imaging 1995: Physics of Medical Imaging
Richard L. Van Metter; Jacob Beutel, Editor(s)

© SPIE. Terms of Use
Back to Top