Share Email Print
cover

Proceedings Paper

Optical fiber power measurement using different transfer standards
Author(s): Nan Xu; Haiyong Gan; Jianwei Li; Zhixin Zhang; Jian Li
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

A comparison of different transfer standard optical fiber power detectors is present. Traceable to cryogenic radiometer, these planar, focus-planar and trap detectors have different characteristics during the optical fiber power values transfer because of the different input angles or fiber connectors. For different types of fibers and fiber connectors, a new trap detector is capable for the optical fiber power measurement, which has very little sensitivity for a variety of input conditions. Comparison of fiber power measurement using a planar and a trap detector is present by employing a three-lens method. A good agreement between the two types of detectors shows the feasibility of fiber power transfer using planar detectors.

Paper Details

Date Published: 13 November 2014
PDF: 6 pages
Proc. SPIE 9276, Optical Metrology and Inspection for Industrial Applications III, 927615 (13 November 2014); doi: 10.1117/12.2073687
Show Author Affiliations
Nan Xu, National Institute of Metrology (China)
Haiyong Gan, National Institute of Metrology (China)
Jianwei Li, National Institute of Metrology (China)
Zhixin Zhang, National Institute of Metrology (China)
Jian Li, National Institute of Metrology (China)


Published in SPIE Proceedings Vol. 9276:
Optical Metrology and Inspection for Industrial Applications III
Sen Han; Toru Yoshizawa; Song Zhang, Editor(s)

© SPIE. Terms of Use