A method of reducing measuring time of temperature measurements of biological objects based on preheating the resistance temperature detector (RTD) up to the temperature close to the temperature to be measured, is proposed. It has been found that at the same measuring time, the preheating allows to decrease the measurement error by a factor of 5 to 45 over the temperature range of 35-41°С. The measurement time is reduced by 1.6-4 times over this range, keeping the same value of the measurement error.

Paper Details

Date Published: 19 August 2014
PDF: 6 pages
Proc. SPIE 9291, 13th International Scientific Conference on Optical Sensors and Electronic Sensors, 92910A (19 August 2014); doi: 10.1117/12.2070167