Share Email Print
cover

Proceedings Paper

Laser Doppler vibrometry of the human middle ear: the technical point of view
Author(s): Hans-Jochen Foth; Matthias Brenner; Norbert Stasche; Karl Hoermann
Format Member Price Non-Member Price
PDF $14.40 $18.00
cover GOOD NEWS! Your organization subscribes to the SPIE Digital Library. You may be able to download this paper for free. Check Access

Paper Abstract

Laser-Doppler-vibrometry is used successfully to detect the motions of the tympanic membrane under normal hearing conditions. For more detailed information about the dynamical behavior of the middle ear, it is necessary to improve the experimental signal-to- noise-ratio, which is mainly determined by the amount of light coupled back into the interferometer. The paper focuses onto two possibilities to yield of this amount light by increasing the solid angle of the optics and increasing the reflectivity of the tympanic membrane.

Paper Details

Date Published: 30 March 1995
PDF: 8 pages
Proc. SPIE 2390, Optical Biophysics, (30 March 1995); doi: 10.1117/12.206001
Show Author Affiliations
Hans-Jochen Foth, Univ. Kaiserslautern (Germany)
Matthias Brenner, Univ. Kaiserslautern (Germany)
Norbert Stasche, Klinikum Kaiserslautern (Germany)
Karl Hoermann, Univ. Hospital Mannheim (Germany)


Published in SPIE Proceedings Vol. 2390:
Optical Biophysics
Halina Podbielska, Editor(s)

© SPIE. Terms of Use