Share Email Print
cover

Proceedings Paper

An integrated fiber and stone basket device for use in Thulium fiber laser lithotripsy
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

The Thulium fiber laser (TFL) is being explored as an alternative laser lithotripter to the Holmium:YAG laser. The TFL's superior near-single mode beam profile enables higher power transmission through smaller fibers with reduced proximal fiber tip damage. Recent studies have also reported that attaching hollow steel tubing to the distal fiber tip decreases fiber degradation and burn-back without compromising stone ablation rates. However, significant stone retropulsion was observed, which increased with pulse rate. In this study, the hollow steel tip fiber design was integrated with a stone basket to minimize stone retropulsion during ablation. A device was constructed consisting of a 100-μm-core, 140-μm-OD silica fiber outfitted with 5-mm-long stainless steel tubing at the distal tip, and integrated with a 1.3-Fr (0.433-mm-OD) disposable nitinol wire basket, to form an overall 1.9-Fr (0.633-mm- OD) integrated device. This compact design may provide several potential advantages including increased flexibility, higher saline irrigation rates through the ureteroscope working channel, and reduced fiber tip degradation compared to separate fiber and stone basket manipulation. TFL pulse energy of 31.5 mJ with 500 μs pulse duration and pulse rate of 500 Hz was delivered through the integrated fiber/basket device in contact with human uric acid stones, ex vivo. TFL stone ablation rates measured 1.5 ± 0.2 mg/s, comparable to 1.7 ± 0.3 mg/s (P > 0.05) using standard bare fiber tips separately with a stone basket. With further development, this device may be useful for minimizing stone retropulsion, thus enabling more efficient TFL lithotripsy at higher pulse rates.

Paper Details

Date Published: 4 March 2014
PDF: 8 pages
Proc. SPIE 8926, Photonic Therapeutics and Diagnostics X, 89261J (4 March 2014); doi: 10.1117/12.2044308
Show Author Affiliations
Christopher R. Wilson, The Univ. of North Carolina at Charlotte (United States)
Thomas C. Hutchens, The Univ. of North Carolina at Charlotte (United States)
Luke A. Hardy, The Univ. of North Carolina at Charlotte (United States)
Pierce B. Irby, Carolinas Medical Ctr. (United States)
Nathaniel M. Fried, The Univ. of North Carolina at Charlotte (United States)
Carolinas Medical Ctr. (United States)
Johns Hopkins Medical Institutions (United States)


Published in SPIE Proceedings Vol. 8926:
Photonic Therapeutics and Diagnostics X
Bernard Choi; Hyun Wook Kang; Brian J. F. Wong; Guillermo J. Tearney; Andreas Mandelis; Nikiforos Kollias; Kenton W. Gregory; Justus F. Ilgner; Haishan Zeng; Laura Marcu, Editor(s)

© SPIE. Terms of Use
Back to Top