Material basis decomposition in the sinogram domain requires accurate knowledge of the forward model in spectral CT. Misspecifications over a certain limit will result in biased estimates and make quantum limited quantitative CT difficult. We present a method whereby users can determine the degree of allowed misspecification error in a spectral CT forward model, and still have quantification errors that are quantum limited.