Photoacoustic correlation spectroscopy for calibration-free absolute quantification of particle concentration
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

Currently, laser fluence calibration is typically required for quantitative measurement of particle concentration in photoacoustic microscopy. In this paper, we present another quantitative approach to measure absolute absorber concentrations by photoacoustic correlation spectroscopy. The proposed method is based on the fact that the Brownian motion induces particle count fluctuation in the detection volume. We first derived a theoretical model for photoacoustic signals and then applied our method to quantitative measurement of different concentrations of various particles. The experimental results agreed well with the predictions from the theoretical model, suggesting that our method can be used for absolute particle concentrations measurement.

Paper Details

Date Published: 3 March 2014
PDF: 6 pages
Proc. SPIE 8943, Photons Plus Ultrasound: Imaging and Sensing 2014, 89431W (3 March 2014); doi: 10.1117/12.2036635
Show Author Affiliations
Yong Zhou, Washington Univ. in St. Louis (United States)
Junjie Yao, Washington Univ. in St. Louis (United States)
Konstantin I. Maslov, Washington Univ. in St. Louis (United States)
Lihong V. Wang, Washington Univ. in St. Louis (United States)


Published in SPIE Proceedings Vol. 8943:
Photons Plus Ultrasound: Imaging and Sensing 2014
Alexander A. Oraevsky; Lihong V. Wang, Editor(s)