Share Email Print
cover

Proceedings Paper

Analysis of sampling error in FTIR
Author(s): Ya Kang
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

In Fourier transform infrared spectrometer (FTIR), the optical path difference of laser metrology is applied to increase spectrum measure precision. However, due to the characteristics of laser single frequency and steady frequency, the uneven speed of moving mirror, circuit delay etc. cause the deviation of sampling point. In this paper, based on the interference theory of spectrometer, according to the Fourier contrary transform and methods of error analysis, the theory model between relative error of spectrum measure and the deviation of sampling point is established. Some simulation computations of the model have been done, the result indicates that, as a theory basis, the model can be applied in analyzing the sampling error of spectrometer and correction algorithm.

Paper Details

Date Published: 4 March 2013
PDF: 6 pages
Proc. SPIE 8761, PIAGENG 2013: Image Processing and Photonics for Agricultural Engineering, 87610N (4 March 2013); doi: 10.1117/12.2019626
Show Author Affiliations
Ya Kang, Chongqing City Management College (China)


Published in SPIE Proceedings Vol. 8761:
PIAGENG 2013: Image Processing and Photonics for Agricultural Engineering
Honghua Tan, Editor(s)

© SPIE. Terms of Use