Share Email Print
cover

Proceedings Paper

Photoacoustic microscopy with 7.6-μm axial resolution
Author(s): Chi Zhang; Konstantin Maslov; Junjie Yao; Lihong V. Wang
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

The axial resolution of photoacoustic microscopy (PAM) is much lower than its lateral resolution, which resolves down to the submicron level. Here we achieved so far the highest axial resolution of 7.6 μm by using a commercial 125 MHz ultrasonic transducer for signal detection, followed by the Wiener deconvolution for signal processing. The axial resolution was validated by imaging two layers of red ink in a wedge shape. Melanoma cells were imaged ex vivo with high axial resolution. Compared with a PAM system with a 50 MHz ultrasonic transducer, our high-axial-resolution PAM system resolved the blood vessels in mouse ears in vivo much more clearly in the depth direction.

Paper Details

Date Published: 4 March 2013
PDF: 6 pages
Proc. SPIE 8581, Photons Plus Ultrasound: Imaging and Sensing 2013, 85812W (4 March 2013); doi: 10.1117/12.2003248
Show Author Affiliations
Chi Zhang, Washington Univ. in St. Louis (United States)
Konstantin Maslov, Washington Univ. in St. Louis (United States)
Junjie Yao, Washington Univ. in St. Louis (United States)
Lihong V. Wang, Washington Univ. in St. Louis (United States)


Published in SPIE Proceedings Vol. 8581:
Photons Plus Ultrasound: Imaging and Sensing 2013
Alexander A. Oraevsky; Lihong V. Wang, Editor(s)

© SPIE. Terms of Use