An interferometric type fiber-optic pressure sensor is constructed by attaching a 125 micrometers single-mode fiber inside a 1.4 mm outer diameter miniature tubing capped with a 2.5 micrometers diaphragm. The optical fiber looks at the diaphragm displacement caused by pressure. The gap between the fiber end and the diaphragm forms a Fabry-Perot interferometer. The number of optical fringes at the output is directly related to the pressure acting on the diaphragm. A second sensor is constructed using a 925 micrometers outer diameter micro bellows. The fiber- optic sensor data compares well with a strain gauge type pressure transducer.

Paper Details

Date Published: 28 July 1994
PDF: 7 pages
Proc. SPIE 2131, Biomedical Fiber Optic Instrumentation, (28 July 1994); doi: 10.1117/12.180752
Show Author Affiliations
Nadarajah Narendran, Mechanical Technology Inc. (United States)
L. Hoogenboom, Mechanical Technology Inc. (United States)
S. Bardsley, Mechanical Technology Inc. (United States)