Multifocal ophthalmic lens testing
Author(s): Rong Sun; Valdemar Portney
Format Member Price Non-Member Price
PDF $14.40 $18.00
cover GOOD NEWS! Your organization subscribes to the SPIE Digital Library. You may be able to download this paper for free. Check Access

Paper Abstract

Optical characteristics of the multifocal intraocular lens (IOL) are different from regular monofocal IOLs. We have developed a test method and procedure to measure completed optical performance of the AMO ArrayTM multifocal IOL and to control its optical quality. The measured parameters include effective focal length, aqueous diopter power, add power, depth-of-focus, resolution efficiency, contrast modulation for far and near image, subjective quality factor variance , and through focus response. Automated Inspection Station was developed to perform the measurements. In addition, surface characteristics of the multifocal IOL are measured by a special custom-built interferometer (Wavy Optical Tester).

Paper Details

Date Published: 26 May 1994
PDF: 6 pages
Proc. SPIE 2127, Ophthalmic Lens Design and Fabrication II, (26 May 1994); doi: 10.1117/12.176828
Show Author Affiliations
Rong Sun, Allergan Medical Optics (United States)
Valdemar Portney, Allergan Medical Optics (United States)


Published in SPIE Proceedings Vol. 2127:
Ophthalmic Lens Design and Fabrication II