Share Email Print
cover

Proceedings Paper

Heat effect of pulsed Er:YAG laser radiation
Author(s): Raimund Hibst; Ulrich Keller
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

Pulsed Er:YAG laser radiation has been found to be effective for dental enamel and dentin removal. Damage to the surrounding hard tissue is little, but before testing the Er:YAG laser clinically for the preparation of cavities, possible effects on the soft tissue of the pulp must be known. In order to estimate pulp damage , temperature rise in dentin caused by the laser radiation was measured by a thermocouple. Additionally, temperature distributions were observed by means of a thermal imaging system. The heat effect of a single Er:YAG laser pulse is little and limited to the vicinity of the impact side. Because heat energy is added with each additional pulse , the temperature distribution depends not only on the radiant energy, but also on the number of pulses and the repetition rate. Both irradiation conditions can be found , making irreversible pulp damage either likely or unlikely. The experimental observations can be explained qualitatively by a simple model of the ablation process.

Paper Details

Date Published: 1 June 1990
PDF: 8 pages
Proc. SPIE 1200, Laser Surgery: Advanced Characterization, Therapeutics, and Systems II, (1 June 1990); doi: 10.1117/12.17481
Show Author Affiliations
Raimund Hibst, Univ. of Ulm (Germany)
Ulrich Keller, Univ. of Ulm (Germany)


Published in SPIE Proceedings Vol. 1200:
Laser Surgery: Advanced Characterization, Therapeutics, and Systems II
Stephen N. Joffe; Kazuhiko Atsumi, Editor(s)

© SPIE. Terms of Use
Back to Top