Paper Abstract

Many techniques exist for statistically estimating measurement uncertainties in radiometric instruments. One method, the fidelity interval, is addressed here and applied to simulated calibration data for the multi-angle imaging spectroradiometer (MISR). The fidelity interval is the region within which the true radiance is expected to lie with a given probability and thus provides a measure of the calibration uncertainty. Calibration data are reduced using a least- squares approach to provide the gain and offset coefficients used in data reduction. It is from these calibration measurements that the fidelity interval is estimated. The technique allows us to optimize our calibration test plan through simulations which vary the number of radiometric levels, the radiometric range available for testing, and the number of data repetitions.

Paper Details

Date Published: 25 August 1993
PDF: 12 pages
Proc. SPIE 1939, Sensor Systems for the Early Earth Observing System Platforms, (25 August 1993); doi: 10.1117/12.152838
Show Author Affiliations
Nadine Lu Chrien, Jet Propulsion Lab. (United States)
Carol J. Bruegge, Jet Propulsion Lab. (United States)
Bruce R. Barkstrom, NASA Langley Research Ctr. (United States)