Potential for unnecessary patient exposure from the use of storage phosphor imaging systems
Author(s): Matthew T. Freedman; Einar V. Pe; Seong Ki Mun; Shih-Chung Benedict Lo; Martha C. Nelson
Format Member Price Non-Member Price
PDF $14.40 $18.00
cover GOOD NEWS! Your organization subscribes to the SPIE Digital Library. You may be able to download this paper for free. Check Access

Paper Abstract

It is possible to use storage phosphor radiography (SR) devices in a manner that results in excess exposure to the patient without the operators knowledge. Because these SR systems have an automatic correction for the final optical density (OD) of the image, the technologist and radiologist will not be able to use excessive blackness of the image as a sign of overexposure. Tests reported here demonstrate that it is possible to obtain images of a chest phantom that appear acceptable with a 32 times difference in exposure (maximal exposure .86 R). It is possible to obtain exposures of a pelvis phantom that appear acceptable up to the tube limit of our machine (4.8 R). Tests of the Fuji AC-1 demonstrate that it will accept a much wider range of exposures than the AGFA ADC prototype which permits only a 7 times difference in exposure before the image is degraded.

Paper Details

Date Published: 30 June 1993
PDF: 8 pages
Proc. SPIE 1897, Medical Imaging 1993: Image Capture, Formatting, and Display, (30 June 1993); doi: 10.1117/12.146998
Show Author Affiliations
Matthew T. Freedman, Georgetown Univ. Medical Ctr. (United States)
Einar V. Pe, Georgetown Univ. Medical Ctr. (United States)
Seong Ki Mun, Georgetown Univ. Medical Ctr. (United States)
Shih-Chung Benedict Lo, Georgetown Univ. Medical Ctr. (United States)
Martha C. Nelson, Georgetown Univ. Medical Ctr. (United States)


Published in SPIE Proceedings Vol. 1897:
Medical Imaging 1993: Image Capture, Formatting, and Display
Yongmin Kim, Editor(s)

© SPIE. Terms of Use
Back to Top