Share Email Print
cover

Proceedings Paper

Design and initial performance evaluation of a portable short-wave infrared spectroradiometer
Author(s): Mark W. Smith
Format Member Price Non-Member Price
PDF $14.40 $18.00

Paper Abstract

The design, initial calibration, and performance evaluations of a portable short wave infrared (SWIR) spectroradiometer are described. The spectroradiometer covers the range from 1.1 to 2.5 microns with a spectral resolution that may be varied from less than 10 nm to more than 100 nm. A single spectrum is acquired in about 2 sec. The SNR is about 230 at a wavelength of 2.2 microns for a Lambertian surface of 90-percent reflectance illuminated by the sun at normal incidence with 14.8-nm resolution, a 25 C background temperature, and no atmospheric attenuation. FOV-defining optics are coupled by a flexible fiber-optics bundle to the spectroradiometer, which consists of a concave holographic diffraction grating with a flat focal field imaged onto a 1024-element platinum silicide linear-array detector.

Paper Details

Date Published: 5 January 1993
PDF: 17 pages
Proc. SPIE 1762, Infrared Technology XVIII, (5 January 1993); doi: 10.1117/12.138951
Show Author Affiliations
Mark W. Smith, Optical Sciences Ctr./Univ. of Arizona (United States)


Published in SPIE Proceedings Vol. 1762:
Infrared Technology XVIII
Bjorn F. Andresen; Freeman D. Shepherd, Editor(s)

© SPIE. Terms of Use