Design and initial performance evaluation of a portable short-wave infrared spectroradiometer
Author(s): Mark W. Smith
Format Member Price Non-Member Price
PDF $14.40 $18.00
cover GOOD NEWS! Your organization subscribes to the SPIE Digital Library. You may be able to download this paper for free. Check Access

Paper Abstract

The design, initial calibration, and performance evaluations of a portable short wave infrared (SWIR) spectroradiometer are described. The spectroradiometer covers the range from 1.1 to 2.5 microns with a spectral resolution that may be varied from less than 10 nm to more than 100 nm. A single spectrum is acquired in about 2 sec. The SNR is about 230 at a wavelength of 2.2 microns for a Lambertian surface of 90-percent reflectance illuminated by the sun at normal incidence with 14.8-nm resolution, a 25 C background temperature, and no atmospheric attenuation. FOV-defining optics are coupled by a flexible fiber-optics bundle to the spectroradiometer, which consists of a concave holographic diffraction grating with a flat focal field imaged onto a 1024-element platinum silicide linear-array detector.

Paper Details

Date Published: 5 January 1993
PDF: 17 pages
Proc. SPIE 1762, Infrared Technology XVIII, (5 January 1993); doi: 10.1117/12.138951
Show Author Affiliations
Mark W. Smith, Optical Sciences Ctr./Univ. of Arizona (United States)


Published in SPIE Proceedings Vol. 1762:
Infrared Technology XVIII
Bjorn F. Andresen; Freeman D. Shepherd, Editor(s)

© SPIE. Terms of Use