A frequency scanning beam steering technique using an alternating polarity transducer array is investigated for presentation of B-mode ultrasound images for front viewing intracavitary applications. Arrays of alternating polarity PZT elements were computer simulated and fabricated to characterize their beam steering and resolution capabilities. A spherical lens was mounted in front of the array to improve focusing. A personal computer based system was used to test and generate sector images of wire targets using these arrays.

Paper Details

Date Published: 5 November 1992
PDF: 4 pages
Proc. SPIE 1733, New Developments in Ultrasonic Transducers and Transducer Systems, (5 November 1992); doi: 10.1117/12.130592
Show Author Affiliations
Narendra T. Sanghvi, Indiana Univ. School of Medicine (United States)
Ryuichi Shinomura, Hitachi Central Research Lab. (Japan)
C. Nakaya, Hitachi Central Research Lab. (Japan)
K. Katakura, Hitachi Central Research Lab. (Japan)