Paper Abstract

An optical multichannel analyzer was used to develop a reflectance radiometer for measurement of medical diagnostic samples. Sample areas were small, typically 5 mm x 5 mm. Spectral and temporal resolution of the reflectance measurements were 0.5 nm and 17 ms, respectively. Reflectance drifts were found to be less than 0.004 over 2 h with accuracy better than 0.012 compared with a reference method. Flexible instrument design allows use of many different illumination and detection geometries. Applications include performance evaluations of chemical formulations and readhead designs.

Paper Details

Date Published: 1 March 1989
PDF: 6 pages
Opt. Eng. 28(3) 283255 doi: 10.1117/12.7976942
Published in: Optical Engineering Volume 28, Issue 3
Show Author Affiliations
Albert Brunsting, Miles Inc. (United States)
Andrew Dosmann, Miles Inc. (United States)
Gerald Shaffer, Miles Inc. (United States)