Share Email Print
cover

Optical Engineering • Open Access

Correcting for focal-plane-array temperature dependence in microbolometer infrared cameras lacking thermal stabilization

Paper Abstract

Advances in microbolometer detectors have led to the development of infrared cameras that operate without active temperature stabilization. The response of these cameras varies with the temperature of the camera’s focal plane array (FPA). This paper describes a method for stabilizing the camera’s response through software processing. This stabilization is based on the difference between the camera’s response at a measured temperature and at a reference temperature. This paper presents the mathematical basis for such a correction and demonstrates the resulting accuracy when applied to a commercially available long-wave infrared camera. The stabilized camera was then radiometrically calibrated so that the digital response from the camera could be related to the radiance or temperature of objects in the scene. For FPA temperature deviations within ±7.2°C changing by 0.5°C/min , this method produced a camera calibration with spatial-temporal rms variability of 0.21°C, yielding a total calibration uncertainty of 0.38°C limited primarily by the 0.32°C uncertainty in the blackbody source emissivity and temperature.

Paper Details

Date Published: 7 January 2013
PDF: 8 pages
Opt. Eng. 52(6) 061304 doi: 10.1117/1.OE.52.6.061304
Published in: Optical Engineering Volume 52, Issue 6
Show Author Affiliations
Paul W. Nugent, Montana State Univ. (United States)
Joseph A. Shaw, Montana State Univ. (United States)
Nathan J. Pust, Montana State Univ. (United States)


© SPIE. Terms of Use
Back to Top