Share Email Print

Journal of Biomedical Optics

Aberration correction during real time in vivo imaging of bone marrow with sensorless adaptive optics confocal microscope
Author(s): Zhibin Wang; Dan Wei; Ling Wei; Yi He; Guohua Shi; Xunbin Wei; Yudong Zhang
Format Member Price Non-Member Price
PDF $20.00 $25.00
cover GOOD NEWS! Your organization subscribes to the SPIE Digital Library. You may be able to download this paper for free. Check Access

Paper Abstract

We have demonstrated adaptive correction of specimen-induced aberration during in vivo imaging of mouse bone marrow vasculature with confocal fluorescence microscopy. Adaptive optics system was completed with wavefront sensorless correction scheme based on stochastic parallel gradient descent algorithm. Using image sharpness as the optimization metric, aberration correction was performed based upon Zernike polynomial modes. The experimental results revealed the improved signal and resolution leading to a substantially enhanced image contrast with aberration correction. The image quality of vessels at 38- and 75-μm depth increased three times and two times, respectively. The corrections allowed us to detect clearer bone marrow vasculature structures at greater contrast and improve the signal-to-noise ratio.

Paper Details

Date Published: 12 August 2014
PDF: 7 pages
J. Biomed. Opt. 19(8) 086009 doi: 10.1117/1.JBO.19.8.086009
Published in: Journal of Biomedical Optics Volume 19, Issue 8
Show Author Affiliations
Zhibin Wang, Institute of Optics and Electronics (China)
Univ. of Chinese Academy of Sciences (China)
Dan Wei, Shanghai Jiao Tong Univ. (China)
Ling Wei, Institute of Optics and Electronics (China)
Yi He, Institute of Optics and Electronics (China)
Guohua Shi, Institute of Optics and Electronics (China)
Xunbin Wei, Shanghai Jiao Tong Univ. (China)
Yudong Zhang, Institute of Optics and Electronics (China)

© SPIE. Terms of Use
Back to Top