Share Email Print
cover

Optical Engineering

Subnanosecond optical gating using coax cable input microchannel plate image intensifier
Author(s): Paul Hoess; Karlheinz Fleder; Joachim M. K. Ehrhardt
Format Member Price Non-Member Price
PDF $20.00 $25.00

Paper Abstract

Proximity focused microchannel plate image intensifies (MCPIIs) with a mesh underlay photocathode are analyzed for their iris- ing time, which is found to be of the order of 650 ps. This is much longer than previously reported but is finally explained by the mesh thickness. The reduction of this effect requires further investigation. A newly introduced impedance match has real broadband characteristics, and the irising is fully caused by other effects. The minimum gate time observable was clearly below 1 ns. The mesh underlay MCPII did not open to its full diameter at the shortest applied times. The long laser diode pulse and a too flat voltage slope of the driving generator prevented exact results for subnanosecond timing. Continued development of the system is underway.

Paper Details

Date Published: 1 August 1998
PDF: 4 pages
Opt. Eng. 37(8) doi: 10.1117/1.601734
Published in: Optical Engineering Volume 37, Issue 8
Show Author Affiliations
Paul Hoess, Stanford Computer Optics, Inc. (Germany)
Karlheinz Fleder, Stanford Computer Optics, Inc. (Germany)
Joachim M. K. Ehrhardt, Proxitronic Funk GmbH & Co. KG (Germany)


© SPIE. Terms of Use