Optical Engineering

Interferometer with added flexibility in its use
Author(s): Felix G. P. Peeters
Format Member Price Non-Member Price
PDF $20.00 $25.00
cover GOOD NEWS! Your organization subscribes to the SPIE Digital Library. You may be able to download this paper for free. Check Access

Paper Abstract

In most length-measuring interferometers, the measuring retroreflector must move along a straight line. Only small deviations from this straight line are permitted; otherwise the interference signal becomes too weak. The interferometer can be made immune to movements perpendicular to the straight line by adding a plane mirror to it. The problem of alignment of the extra mirror, which is very critical in existing schemes, is solved by reflecting the reference beam in this plane mirror. However, the perpendicular movements partly alter the course of the optical path. The resulting loss in accuracy when using standard optical elements can be kept below 0.5 µm. With error compensation, the loss in accuracy can be kept below 0.1 µm. For interferometric measurement applications that do not require the highest attainable precision, this is acceptable. If the lateral position of the return beam is restored just prior to each measurement, the accuracy of this optical scheme is comparable to that of a standard one.

Paper Details

Date Published: 1 July 1996
PDF: 4 pages
Opt. Eng. 35(7) doi: 10.1117/1.600629
Published in: Optical Engineering Volume 35, Issue 7
Show Author Affiliations
Felix G. P. Peeters, Eindhoven Univ. of Technology (Netherlands)